Norlab

Your partner in Northern Europe for innovative laboratory instruments

Biotage Extrahera precision and accuracy performance part 2

Resource type: 
Technical note
Author(s): 
Biotage
Format: 
pdf
Date of creation: 
3 December 2015
Library code: 
10484

This document details an example of the pipetting precision and accuracy data for four specified volumes across 24 (4 x 6) positions.

Experimental design

  • Precision and accuracy were tested by separately measuring the volume of 24 transfers of 50 uL, 100 uL, 500 uL and 1000 uL
  • Water was dispensed sequentially (4 channels simultaneously) starting with column 1 (positions A1-D1) and ending with column 6 (positions A6-D6)
  • Performance was assessed by measuring the weights of 24 collection tubes before and after the addition of purified water
  • Volume data was generated for each position using temperature corrected density values

Because of the low volumes and weights involved, correction factors for temperature and evaporation rate were applied.